UV–Cabinets for PCR operations (UVC/T-AR,
UVC/T-M-AR, UVT-B-AR and UVT-S-AR) are
designed for clean operations with DNA samples.
They provide protection against contamination.
All models are bench-top type, made of metal
framework, glass (or plexiglas) walls and working
surface painted with powder enamel or made of
stainless steel
UV–Cabinets are equipped with an open UV
lamp installed in the upper hood. UV-radiation
from the open lamps disinfects the working area
inactivating DNA/RNA fragments during 15–30 min
of exposure. A digital timer controls duration of
the direct UV irradiation. |
|
UV–Cabinets are equipped with a flow-type
bactericidal UV cleaner–recirculator AR, which
provides constant decontamination inside the
cabinet during operation. They are recommended
for operations with DNA/RNA amplicons.
UV cleaner–recirculator AR consists of a UV lamp,
a fan and dust filters organized in a special body so
that a user working with a UV–Cabinet is protected
against UV light. Recirculator increases the
maximum density of UV light making it sufficiently
effective for DNA/RNA inactivation. The UV–recirculator
processes 100 UV–Cabinet volumes per hour,
creating permanent asept |
Walls material |
Rear: stainless steel Sides and front: glass |
Working surface material |
Stainless steel |
Open UV-lamp |
1 x 25W |
UV radiation level |
15 mW / cm2 / sec |
Radiation type |
UV (λ = 253.7 nm), ozone-free |
Digital time setting of direct UV exposure |
0–24 hrs / non–stop (increment 1 min) |
UV-recirculator |
1x25W (efficiency >99% per 1 hour) |
Daylight lamp (for working area illumination) |
1 x TLD-15W |
Thickness of side panels |
4 mm |
Thickness of upper front panel |
8 mm |
|
Thickness of screen |
4 mm |
Optical transmission |
95% |
UV protection |
>96%, UV-protection film |
Working area |
650 × 475 mm |
Safety |
Automatic open UV-lamp switch off when screen is open |
Power outlets inside the unit |
Inlet for power cords |
Nominal operating voltage |
230 V, 50 Hz or 120 V, 60 Hz |
Power consumption (230V / 120 V) |
253 VA (1.2 A) / 372 VA (2 A) |
Overall dimensions |
690 × 535 × 555 mm |
Optional table |
T-4 (800 × 600 × 750 mm) |
|
Advantages of Biosan UV–Cabinets:
Ozone free high density UV decontamination
Long living UV lamps (8,000 hours average)
Automatic switch off of UV-lamps when the protective
screen is opened
Bactericidal flow-type recirculator providing permanent
decontamination inside UV –cabinet during
operation
PCR CABINET BIOSAN
|
Shockproof glass walls
Low noise, low energy consumption
Tables for installation of UV–Cabinets
UV–Cabinets with the bactericidal
UV cleaner–recirculator AR is the patented Biosan
PCR CABINETS BIOSAN
|
|