MAIN PAGE  
OUR PRODUCTION
DISTRIBUTOR

 

MEDICAL EQUIPMENT
LABORATORY CONSUMABLE
PCR CABINETS
UV–Cabinets for PCR operations (UVC/T-AR, UVC/T-M-AR, UVT-B-AR and UVT-S-AR) are designed for clean operations with DNA samples.
They provide protection against contamination.
All models are bench-top type, made of metal framework, glass (or plexiglas) walls and working surface painted with powder enamel or made of stainless steel
UV–Cabinets are equipped with an open UV lamp installed in the upper hood. UV-radiation from the open lamps disinfects the working area inactivating DNA/RNA fragments during 15–30 min
of exposure. A digital timer controls duration of the direct UV irradiation.
 
UV–Cabinets are equipped with a flow-type bactericidal UV cleaner–recirculator AR, which provides constant decontamination inside the cabinet during operation. They are recommended for operations with DNA/RNA amplicons. UV cleaner–recirculator AR consists of a UV lamp, a fan and dust filters organized in a special body so
that a user working with a UV–Cabinet is protected
against UV light. Recirculator increases the maximum density of UV light making it sufficiently effective for DNA/RNA inactivation. The UV–recirculator processes 100 UV–Cabinet volumes per hour, creating permanent asept
Walls material
Rear: stainless steel Sides and front: glass
Working surface material
Stainless steel
Open UV-lamp
1 x 25W
UV radiation level
15 mW / cm2 / sec
Radiation type
UV (λ = 253.7 nm), ozone-free
Digital time setting of direct UV exposure
0–24 hrs / non–stop (increment 1 min)
UV-recirculator
1x25W (efficiency >99% per 1 hour)
Daylight lamp (for working area illumination)
1 x TLD-15W
Thickness of side panels
4 mm
Thickness of upper front panel
8 mm
Thickness of screen
4 mm
Optical transmission
95%
UV protection
>96%, UV-protection film
Working area
650 × 475 mm
Safety
Automatic open UV-lamp switch off when screen is open
Power outlets inside the unit
Inlet for power cords
Nominal operating voltage
230 V, 50 Hz or 120 V, 60 Hz
Power consumption (230V / 120 V)
253 VA (1.2 A) / 372 VA (2 A)
Overall dimensions
690 × 535 × 555 mm
Optional table
T-4 (800 × 600 × 750 mm)

Advantages of Biosan UV–Cabinets:
Ozone free high density UV decontamination
Long living UV lamps (8,000 hours average)
Automatic switch off of UV-lamps when the protective screen is opened
Bactericidal flow-type recirculator providing permanent decontamination inside UV –cabinet during operation

PCR CABINET BIOSAN

Shockproof glass walls
Low noise, low energy consumption
Tables for installation of UV–Cabinets
UV–Cabinets with the bactericidal
UV cleaner–recirculator AR is the patented Biosan

PCR CABINETS BIOSAN